other Major Functional Component connected to the LAN.
b. If any master station fails, the dome lights and zone lights shall
continue to indicate calls from patient care rooms.
c. At a minimum, a failure alarm shall be automatically initiated by a
total failure of the NCAV System, by failure of any Major Functional
Component of the system, by failure of any power supply, by failure of any
supervised circuit, and by failure of any circuit or component monitored by
built-in fault diagnostics.
d. If the installed system includes any other built-in fault diagnostics
designed in by the Manufacturer, there shall be automatic failure alarm
indications resulting from these built-in fault diagnostics.
e. Failure alarms shall be annunciated locally on the individual Major
Functional Component, communicated to the Maintenance Workstation, and
transmitted to pagers carried by the medical facility maintenance staff.
1.4.4
The Contractor shall submit Coordination Drawings as specified in the
voice intercom network diagram, riser diagrams, wiring and schematic
diagrams, run sheets including number of conductors and wire number (ID),
custom assembly details, and installation details.
b. Riser diagrams shall indicate the identification number (ID) for all
zone lights as shown on the plan drawings.
c. Installation details shall indicate layout and mounting of equipment,
equipment relationship to other parts of the work, including clearances
required for maintenance and operation, and plan and elevation details that
indicate the exact and totally coordinated physical location and size of
each individual item of equipment.
d. Details for the custom assembly of equipment shall indicate the
assembly configuration, elevations and dimensions. Typical custom assembly
details include equipment panels, and equipment mounted in a rack or
cabinet.
1.5
QUALITY ASSURANCE
1.5.1
Qualifications
1.5.1.1
General Qualification Requirements
a.
The NCAV System Contractor, Installer and Manufacturer shall each have
the minimum qualifications specified, related to the type of system
specified for this project.
b.
The Government reserves the right to accept or reject the system
Contractor, Installer or Manufacturer based upon qualifications and ability
to conform to specified requirements of this Section. System Contractors,
Installers and Manufacturers that do not have the specified qualifications
will not be acceptable and shall not be allowed to perform the work of this
Section.
SECTION 16731A
Page 31