1.4.2.9
Servers
a. One or more servers, connected to the system LAN, shall be provided as
required by the system application design for the following functions:
(1) Call Logging Server. Provides the central database for all call
logging data from all NC Subsystems, and the data interface with the
medical facility Information System LAN to allow medical facility
Information System terminals and computers to access call logging data and
reports.
(2) Information System Server. Provides the data interface with the
medical facility Information System for the downloading of patient ADT data
to all NC Subsystems.
(3) Radio Page Server. Provides the data interface with the Radio
Paging System to communicate calls from throughout the system for
transmission to alphanumeric pagers that are carried by the medical
facility on-duty staff.
(4) Wireless Telephone Server. Provides the data interface with the
Wireless Telephone System to communicate calls from throughout the system
for transmission to the alphanumeric text messaging feature of wireless
phones that are carried by the medical facility on-duty staff.
required for each specified interface function.
c. The location of servers shall be as required by the system application
design.
1.4.2.10
Central Master Station - Future Option
The system shall have the capability and system architecture to allow the
future addition of an optional central master station that can receive and
answer calls from any combination of all patient care areas. The future
implementation of this option shall not require any changes in the system
architecture nor require the replacement of any installed system equipment.
1.4.3
Performance Requirements
1.4.3.1
NC Subsystems Functions and Features
a. Communication of calls for assistance and information, medical device
alarms, and patient safety and security alarms.
(1) Patient safety alarm calls are from the interfaced Bed Exit
System.
(2) Security alarm calls are from the interfaced Infant Protection
System, Patient Wandering System, and the medical facility Intrusion
Detection System.
(3) Call features shall be as specified herein.
b. Within each patient bedroom, operate and listen to the sound from the
patient TV Set via a hardwired interface with the TV Set. This function
shall utilize the control and speaker features of both the connected pillow
speaker and bed SideCom simultaneously. The TV sound shall be muted during
SECTION 16731A
Page 20