a. [Automatic controls] [Microcomputer controls with cycle
documentation via strip recorder, for each cycle. Microcomputer
controls must allow for programming chamber aeration cycle at same
temperature as sterilizing cycle, adjustable from 1 to 59 minutes, 59
seconds. Controls designed to facilitate mounting at location up to
15,000 mm 50 feet from sterilizer].
b.
[Jacket condensate to return] [Condensate to waste].
c.
Manual or power operating door.
d.
Electrical characteristics:
(1) Vacuum pump: [120] [200] [230] [460] volt, 60 Hz, [single]
[three] phase.
(2) Controls and similar equipment: 120 volt, 60 Hz, single phase.
2.8.1
Pack Sterilization
**************************************************************************
NOTE: Do not return condensate from units located
so far from steam boiler that it is uneconomical.
Only jacket steam condensate is reusable.
**************************************************************************
Sterilizer shall be electrically heated, gravity air removal type, single
door unit, recess mounted through one wall, with manually operating door,
either hinged or sliding. Unit suitable for sterilization of packs,
utilizing saturated steam or sterilizing agent. Challenge packs shall
conform to AAMI ST8. The following shall be included:
a. [Automatic controls] [Microcomputer controls with cycle
documentation via strip recorder, for each cycle. Microcomputer
controls must allow for programming chamber aeration cycle at same
temperature as sterilizing cycle, adjustable from one to 59 minutes, 59
seconds. Controls should be designed to facilitate mounting at
location up to 15,000 mm 50 feet from sterilizer].
b.
c.
Electrical characteristics:
[120] [200] [230] [460] volt, 60 Hz [single]
[three] phase.
(2) Controls:
120 volt, 60 Hz, single phase.
d. [Materials handling accessories, including removable rack with
three full-length removable shelves] [Materials handling accessories,
including loading car and transfer carriage].
2.8.2
Solution Sterilization
Solution performance test shall conform to AAMI ST8.