]2.2
AUDIO EQUIPMENT
Audio functions of nurse call equipment shall be built-in features. Each
control station shall contain an amplifier with sufficient output to
address all patient stations within the control area, simultaneously.
Provide two-way hands-free communications to the control station from each
patient, staff, or duty station. Patient, or staff, shall be able to
converse with the control station attendant without moving, without
directing toward the transmitter, without using controls, and without
raising the voice above a normal speaking level. Adjustable volume
controls shall be contained in the control station.
2.3
SIGNALS
**************************************************************************
NOTE: Verify the priority level of functions and
associated signals in use at the facility. Modify
these requirements as necessary.
**************************************************************************
Priority levels and dome light indications shall be as recommended in IEEE
Std 602 [except that code call signals shall be flashing red and Code Blue
signals shall be flashing blue].
2.4
WIRE/CABLE
Provide interconnecting wire/cable as accessory equipment and comply with
the same standards as the equipment with which used.
2.5
RACEWAYS
Provide metal raceways conforming to Section 16402 INTERIOR DISTRIBUTION
SYSTEM. Fittings for EMT shall be compression type with ferrule and gland
nut, not set screw.
PART 3
EXECUTION
3.1
INSTALLATION
**************************************************************************
NOTE: This guide was prepared for use in preparing
construction. The guide should be modified as
appropriate if used for a remodeling or retrofit
type project. In retrofit projects, the designer
should become familiar with as-built conditions, and
maximize the use of existing conduits and raceway
components to the practical extent. Use extreme
care in retrofit specifications to avoid proprietary
statements. Add specifications to provide for
repair of existing areas that are disturbed by the
retrofit. Show additional drawing details
sufficient to allow the prospective contractor to
bid all aspects of the job.
**************************************************************************
Provide a complete and operational nurse call system, with subsystems, to
satisfy the specified performance. Install equipment and accessory items
to suit manufacturer's instructions and recommendations. Provide insulated
SECTION 16725N