PART 3
EXECUTION
3.1
INSTALLATION
Section 11700 GENERAL REQUIREMENTS FOR MEDICAL AND DENTAL EQUIPMENT, and
NFPA 70. Install lighthead in accordance with the approved installation
drawings and submit installation report for each lighthead.
3.1.1
Wiring Methods
Provide conduit and wiring in accordance with Section 16402 INTERIOR
DISTRIBUTION SYSTEM.
[3.1.1.1
Outlet Box
**************************************************************************
NOTE: Delete this paragraph if lighthead intensity
control is surface mounted.
**************************************************************************
Provide three-gang outlet box 75 mm 3 inches depth for recessed mounted
intensity control. Install box 1525 mm 5 feet from finished floor to
center line of box.
]3.1.2
Surgical Light Transformer
**************************************************************************
of wires shall not exceed 9144 mm 30 feet from
transformer to the lighthead.
**************************************************************************
Provide [mounting bracket] [seismic anchoring] for the transformer. Mount
transformer on the [ceiling] [wall, 1525 mm 5 feet, minimum, from finished
floor].
3.2
FIELD QUALITY CONTROL
3.2.1
Inspection
Examine each item visually for conformance to the requirements of this
section.
3.2.2
Tests
Upon completion of installation, conduct an operating test to demonstrate
that each surgical lighting fixture meets the requirements of this section.
-- End of Section --
SECTION 16553N
Page 9